Terumo Sarns Modular Perfusion System 8000 Heart Lung Machine

Terumo Sarns Modular Perfusion System 8000 Features

The Terumo Sarns Modular Perfusion System 8000 is a flexible, dependable, and cost-efficient modular blood pumping system.The Sarns 8000 is a modular heart-lung machine that is available in a variety of configurations. We offer 4 and 5 pumps units that are available with and without the built-in centrifugal pump. All units have either been factory refurbished or have been well maintained with a complete service record.

  • Modular design
  • Four- or five-pump base
  • Roller or centrifugal pumps
  • Arterial and cardioplegia monitors for temperature, pressures and timers
  • Battery module
  • Electronic venous occluder
  • Pulse mode
  • Communications module for data acquisition

Terumo Sarns Modular Perfusion System 8000 Specifications

View List

Dimensions

  • Telescope pole: 1.30” (3.3 cm) diameter, 6-7 feet (1.8 – 1.9 m) height

4-Pump Unit

  • Height 25.8” (65.5 cm)
  • Width 33.3” (84.5 cm)
  • Depth 26.5” (67.3 cm)
  • Weight 330 lbs (149.5 kg)

5-Pump Unit

  • Height 25.8” (65.5 cm)
  • Width 41.6” (105.6 cm)
  • Depth 26.5” (67.3 cm)
  • Weight 380 lbs (172.1 kg)

Pump / Roller

  • Height 14” (35.6 cm)
  • Width 8.2” (20.8 cm)
  • Depth 19.7” (50 cm)
  • Weight 50 lbs (22.6 kg)

Electrical Power Performance

  • Leakage Current: <100 microamperes for 115 V and 100 V bases (fully loaded);<200 microamperes for 220-240 V bases (fully loaded). The maximum leakage current for the base without pumps is 45 microamperes for all input voltages
  • Ground resistance: 0.1 ohms maximum for all bases
  • Emergency power: 8000 system battery (two 12 volt, sealed gel cell batteries) will power the arterial pump (operating in continuous mode), the monitors, occluder and lamp for 40 minutes; when the arterial pump is operated in pulsatile mode or the cardioplegia pump is also operating on battery power, the battery time available will diminish to 25 minutes
  • The base is tested for high potential: 1200V~ for 100-115 VAC units; 1500V~ for 220-240 VAC units